How to use the disintegration tester


Release Date:

2022-04-18

According to the pharmacopoeia's requirements for disintegration, unless otherwise specified, tablets with measured solubility no longer need to be checked for disintegration time. Disintegration testers are widely used in refining fields, such as health products, refined traditional Chinese medicine, sugar coating, and enteric preparations. How is the disintegration tester used in that experiment?

According to the pharmacopoeia's disintegration requirements, unless otherwise specified, tablets whose solubility is measured no longer require disintegration time testing. Disintegration testers are widely used in the refinement field, such as health products, traditional Chinese medicine refinement, sugar coating, and enteric coatings. How was the disintegration tester used in that experiment?


When using the intelligent disintegration tester for the first time, the disintegration basket should be set high. The steps are as follows: Place the beaker in the water bath; place the basket cylinder upwards on the beaker; log in as an administrator, open the "equipment settings" interface, click "base", and minimize the basket's cross arm; assemble the basket and secure it to the cross arm of the intelligent disintegration tester. The disintegration basket is set high, and the process is complete.


Before starting the disintegration test, we need to set the parameters of the disintegration tester. The setting steps are as follows: Power on; instrument self-test; user login; set water bath temperature and preset rise time; start heating.


Disintegration test process: Pour the experimental solution into the beaker and place it in the water bath; install the hanging basket; the small end of the water level gauge must be inserted vertically into the hanging basket glass tube to ensure that the liquid level in the beaker is no lower than the step height of the water level gauge (Note: The step height of the water level gauge is 15 ± 0.5 mm). Otherwise, the amount of liquid in the beaker; the water bath temperature must be adjusted to the required temperature.


During the experiment, the frosted surface lighting system can be turned on to observe the disintegration of the tablets. For example, if the tablets disintegrate prematurely, the pause button can be clicked to end the experiment early.


Performance characteristics of the disintegration tester


Timing range: Randomly set within 10 hours, display resolution is minutes


Temperature preset range: Room temperature - 45.0 randomly set, display resolution 0.1.


Temperature control accuracy: 0.5


Basket frequency: 30-32 times/minute


Basket lifting distance: 55 mm ± 1 mm


Screen-bottom clearance: 25 mm ± 2 mm


Screen aperture: 2 mm (1 mm aperture optional)


Working power supply: AC 220V ± 10%, 50Hz


Ambient temperature: 5-35℃, relative humidity: 80%


Total power consumption: 550W (including heater power)


Inspection using a disintegration tester clearly cannot serve as an evaluation standard for the in-vivo absorption of all tablets and capsules. Because after the drug dissolves through the disintegration tester, the particles in the net are always between 1.6-2.0 mm, while the drug must be in a solution state to be absorbed by the body, and its particle size is calculated as A, so the disintegration of the disintegration tester is only the initial stage of drug dissolution, and the subsequent continuous dispersion and dissolution process cannot control the decay time limit check. Moreover, the disintegration of solid preparations is affected by many complex factors such as prescription design, preparation, storage process, and in vivo factors. Therefore, the disintegration time limit check cannot objectively reflect the relationship and influence between the drug and excipients. The solubility check includes the disintegration and dissolution process, so studying solubility is more important.